Parts to be Replaced Periodically

HMW-201 Series Exposure Units (for Printed Circuit Board)

Parts Frequency of Replacement Model/Type
Lamp Replace after 5,000 - 7,000 times of flashing (15 sec/shot or less), when both ends are deteriorated in black and the illumination intensity is lowered, or flickering occurs.  (Longer the exposure time is the shorter the lamp life.) IML-3000, IML-4000, IML-5000
Caution:  Using the lamp which steeply exceeded the regulated flashing times may cause the lamp breakage, etc. 
Cooling Water Check the amount of water once a month and supply distilled water if short.  Replace the whole water once a year. Distilled water (Tank Capacity: 5.2 gal)
Ionless Filter The replacement frequency depends upon the frequency of usage.  Normally, replace once every there months. Ionless Filter (WT-0003)
Jacket Tube Replace it when the lamp intensity is still low even after the lamp is replace, or when it is contaminated. Jacket Tube (Inner and Outer)

 

HMW-680 Series Exposure Units (for Printed Circuit Board)

Parts Frequency of Replacement Model/Type
Lamp Replace after 1,000 hour usage, when the both ends are deteriorated in black and the illumination intensity is lowered, or flickering occurs.  SMX-7000H, NMX-7000H, MXA-7000H
Caution:  Using the lamp which steeply exceeded the regulated hours may cause the lamp breakage, etc. 
Cooling Water Check the amount of water once a month and supply distilled water if short.  Replace the whole water once a year. Distilled water (Tank Capacity: 5.2 gal)
Ionless Filter The replacement frequency depends upon the frequency of usage.  Normally, replace once every there months. Ionless Filter (WT-0003)
Jacket Tube Replace it when the lamp intensity is still low even after the lamp is replace, or when it is contaminated. Jacket Tube (Inner and Outer)